THS90Q Industrial Ultra-Low Dew Point Transmitter

Features : Ultra-Low Dew Point Measurement, Fast Response and Low-Moisture Detection
Sensor type : QCM
Dew point scaling range : -100 ... 0 dp°C
Output : 4 ... 20 mA + RS-485 (Standard) / 0 ... 10 V + RS-485 (on request)
Accuracy : -80 ... -20 dp°C: ±2 dp°C / -100 ... -80 dp°C: ±3 dp°C
IP rating : IP54 (Housing) / IP20 (Probe)

THS90Q Industrial Ultra-Low Dew Point Transmitter

Features : Ultra-Low Dew Point Measurement, Fast Response and Low-Moisture Detection
Sensor type : QCM
Dew point scaling range : -100 ... 0 dp°C
Output : 4 ... 20 mA + RS-485 (Standard) / 0 ... 10 V + RS-485 (on request)
Accuracy : -80 ... -20 dp°C: ±2 dp°C / -100 ... -80 dp°C: ±3 dp°C
IP rating : IP54 (Housing) / IP20 (Probe)

Features

Ultra-Low Dew Point Measurement Range
Measurement range from -100 ... 0 dp°C, suitable for compressed air, desiccant drying, PSA nitrogen generation, and ultra-dry processes.
Segmented Accuracy for Ultra-Low Dew Point
Provides corresponding accuracy specifications across low dew point ranges from -100 ... -20 dp°C, helping users monitor dew point changes in ultra-dry processes.
Fast Response and Low-Moisture Detection
The QCM sensing signal provides distinguishable changes under low dew point conditions. With t90 < 30 s at +25 °C, dew point changes can be identified at an early stage.
On-Site Readout and System Integration
Supports display and output of dew point, ppmv, and ppmw values. Equipped with a 2-inch color LCD, 4 ... 20 mA, and RS-485 for convenient on-site reading and system integration.
Stable Measurement Design for Industrial Environments
Enhanced signal processing and EMC protection help reduce the effects of electrical interference and changing environmental conditions on low dew point measurement, improving measurement stability in industrial applications.

Specification

Input

Input type
QCM
Operating range of dew point
-100 ... 0 dp°C

Output

Max. scaling range
Dew point: -80 ... 0 dp°C
Dew point conversion range
Absolute humidity (volume): 0 ... 1000 ppm/v
Absolute humidity (weight): 0 ... 634 ppm/w
Output signal
4 ... 20 mA + RS-485 (Standard)
0 ... 10 V + RS-485 (on request)
Signal connection
3-wire
Linear accuracy
-80 ... -20 dp°C: ±2 dp°C
-100 ... -80 dp°C: ±3 dp°C
Load resistance
Current output: ≤ 500 Ω
Voltage output: ≥ 100 KΩ
Output calibration (ZERO & SPAN) adjustment range
Software
Response time t90 (at +25°C)
< 30 sec

Electrical

Power supply
DC 24 V±15%
Current consumption
DC 24 V / 50 mA
Electrical connection
M12 connector

Environment

Medium
Air
Operating temperature for housing
0 ... +60°C
Operating humidity for housing
0 ... 95%RH (Non-condensing)
Operating temperature for probe
15 ... +50°C
Storage temperature
-25 ... +50°C

Installation

Installation
1/2″G

Protection

IP rating
Housing: IP54 (Probe: IP20)
Electrical protection
■ Reverse polarity ■ Over-voltage ■ Short-circuit

Material

Housing
Aluminum alloy
Probe
Stainless steel SUS304 / Sintered filter SUS316
Weight
250 g

Applications

CDA Compressed Dry Air Systems / PSA Nitrogen Generation and Air Separation Equipment / Desiccant Dryers / Plastic Drying Equipment / Semiconductor and Electronics Manufacturing Drying Systems / Industrial Drying Processes

Note : Design and specifications are subject to change without notice.
Note : Please make sure the product and the device which connect with RS-485 are on common ground, avoid damaged product.