• eYc Ozone Analyzer

eYc Ozone Analyzer

Note:This product only available in Taiwan, China and Hong Kong.
Applicable to measurement and analysis of environmental protection industries such as semiconductor, agriculture, water resources, etc.
UV-D photometric method determination of ozone concentration, traceable to NIST international standards
Measuring range:
Standard type(User-range):1000 ppmv
Standard type(Auto-range):1 ... 100 ppm
Resolution:
Standard type(User-range):Min. 0.1 ppmv at 1000 ppm
Standard type(Auto-range):Min. 0.1 ppbv at 1 ppm
Model : eYc Ozone Analyzer

  • UV-D photometric method determination of ozone concentration, traceable to NIST international standards
  • Wide measuring range / Dynamic full-range measurement mode
  • Quickly switch between measuring range and concentration unit(Six types)
  • LCD, display of concentration, temperature and pressure
  • Automatic temperature and pressure compensation / Built-in aspirator pump
The most commonly used is the principle of Ozone Analyzer is the ultraviolet absorption method, which uses a stable ultraviolet light source to generate ultraviolet rays, and filters other wavelengths of ultraviolet light through a special quartz glass. Only 253.7±1 nm are allowed to pass.
A single stable UV light source can reach the sampling photoelectric transmitter after passing through the highly transparent quartz glass window and the ozone reaction tank.
Through the ratio of the electrical signals(Light intensity) of the sample and the photoelectric sensor that filters ozone-free air, the data can be transmitted after analysis by the management software, we can calculat the ozone concentration.
Ozone generation and treatment technology has a wide range of applications in various industries.
It has the advantages of cleanliness, deodorization, decolorization, sterilization, no residue, and low cost, which is in line with the advantages of green industrial processes.
After the final reaction and half-life of ozone, it will be reduced to oxygen, so there will be no problem of toxic residue.

More potential product articles, welcome to visit the link below
eYc-Success-StoryeYc-FAQ

Type

Standard type(User-range)

Standard type(Auto-range)

Measuring range

1000 ppmv / 1% / 20%v/v / Auto-range

1 ... 100 ppm / User-range

Analog output

Analog output(Optional)

Analog output(Optional)

4 ... 20 mA / DC 0 ... 1 V / DC 1 ... 5 V
(Choose one)

4 ... 20 mA / DC 0 ... 1 V / DC 1 ... 5 V
(Single selection)

Display unit

ppm%mg/m3g/m3
at Auto-range

ppbppmmg/m3、μg/m3
at Auto-range

Heat engine time

5 min

5 min

Resolution

Min. 0.1 ppmv at 1000 ppm

Min. 0.1 ppbv at 1 ppm

Reaction time(T95)

10 sec

40 sec at 500 ppb

Data update speed

0.1 sec ... 100 min

0.1 sec ... 100 min

Data average time

0.1 sec ... 25500 min

0.1 sec ... 25500 min

Data storage

4000 rows of data record
(Max. storage capacity 270 days,
hardware output)
4000 rows of data record
(Max. storage capacity 270 days,
hardware output)

Zero drift

Automatic compensation

Automatic compensation

Accuracy
(Instrument error)

±2%F.S.(Full scale)

Reading±1% at 100 ppb ... 100 ppmv
Provide ISO17025 report traceable to NIST SRP#30

Sample update speed

1 sec ... 24 hr

1 sec ... 24 hr

Detecting unit

254 nm, reduced to 50% after
about 20,000 hours of life

254 nm, reduced to 50% after
about 20,000 hours of life

Sampling &
Calibration flow

1.0 L/min(not included pump)

1.0 ... 1.5 L/min(Built-in pump)

Exterior dimension

L250xW200xH62 mm

L250xW200xH62 mm

Weight

1.5 kg(Include active carbon)

1.5 kg(Include purification device)

Power supply

DC 12 V, 1.5A max.
at AC 100 ... 240 V, 50 / 60 Hz

DC 12 V, 1.5A max.
at AC 100 ... 240 V, 50/60 Hz

Operating Temp. range

0 ... 40°C

0 ... 40°C

Operating pressure range

700 ... 780 mmHg

700 ... 780 mmHg

Digit signal output

USB(Temp. / Pressure / Concentration)

USB(Temp. / Pressure / Concentration)

Warning signal output

LED(Filter clogging/ Light source /
Active carbon / Calibration reminder)

LED(Filter clogging / UV light source /
Ozone purifier / Calibration)


Standard
accessories


User manual

User manual

Silicone tube(0.5 m)

Silicone tube(0.5 m)

B-type USB cable(1 m)

B-type USB cable(1 m)

USB to RS-485 one communication line
(User self-calibration)

None

DC 12 V 1.5 A max. 1
voltage transformer

DC 12 V 1.5 A max. 1
voltage transformer

Verification retrospective report

Calibration retrospective report

Software CD for user self-calibration

None


Optional
accessories


Software CD for user self-calibration and
concentration conversion(Advanced version)

Select two correction range
(Calibration fee is extra)

Float ozone resistant
flow controller(0 ... 2 LPM)

Float ozone resistant
flow controller(0 ... 2 LPM)

None

Software CDRead, self-calibration,
switching and other functions

Pump DC 110 / 220 V 60 Hz

Pump DC 110 / 220 V 60 Hz

None

USB to RS-485 communication cable
Self-calibration, switching and other functions


The specifications contained herein are subject to change without notice and any user of said specifications should verify from the manufacturer that the specifications are currently in effect.
Otherwise, the manufacturer assumes no responsibility for the use of specifications which may have been changed and are no longer in effect.
Mainly used in ozone preparation workshop(Ozone generator / ozone plant building, etc.) / Chemicals / Petroleum / Paper / Textile / Pharmaceutical and essence perfume industry / Waterworks / Sewage / Ozone industry / Industrial process water treatment and ozone disinfection testing in pure water industry / Food and medicine sterilization workshops and other occasions